Phase 1
The plungers of the eight syringes tested were contaminated with cyclophosphamide (Table 1) (mean value, 71.5 ng; range, 3.7-445.7 ng). Cyclophosphamide concentration in the solution was 20 mg/mL, which corresponds to a mean volume of 3.6 nL (0.2-22.3 nL).
Contamination reached 50 ng or more in one of three syringes, and about 5 ng in two of three syringes. No relationship was found between the number of in-and-out movements of the plunger and the quantity of cyclophosphamide on the plunger.
Phase 2
The results of the second phase are shown in Tables 2 and 3. Almost no contamination was found when labelled water was used (A). Contamination remained under 1 nL, even after 10 in-and-out pushes, although a slight increase was noted when the number of plunges increased. The contamination of the plungers was consistently greater with the solution of radiolabelled cyclophosphamide than with the pure radiolabelled solution, regardless of the test or the number of in-and-out pushes. This difference became obvious after the first use of the syringe, whether the operator touched the plunger with gloves or not; however, the total contamination of the plungers was more important after the operator had touched the plunger than otherwise, but this difference disappeared after 10 plunges.
Upper and lower surfaces of the plungers (E1 and E2). The contamination of the upper and lower surfaces of the plungers corresponds to the amount of contaminant that could come into contact with the gloves of operators.
Figure 1. Location of samples from syringe plungers.
Table 1. Amounts and volumes of cyclophosphamide on the plungers of the eight syringes
Table 2. Volumes (nL) of contaminating agents on the plungers of syringes; results of three tests
Bertrand Favier, PharmD
Laurence Gilles, PharmD
Jean Franc¸ois Latour, PharmD
Michel Desage
Francesco Giammarile, PhD
From the Pharmacy Department, Centre Re´gional de Lutte Contre le Cancer Le´on Be´rard, Lyon, France
Phase 1
The plungers of the eight syringes tested were contaminated with cyclophosphamide (Table 1) (mean value, 71.5 ng; range, 3.7-445.7 ng). Cyclophosphamide concentration in the solution was 20 mg/mL, which corresponds to a mean volume of 3.6 nL (0.2-22.3 nL).
Contamination reached 50 ng or more in one of three syringes, and about 5 ng in two of three syringes. No relationship was found between the number of in-and-out movements of the plunger and the quantity of cyclophosphamide on the plunger.
Phase 2
The results of the second phase are shown in Tables 2 and 3. Almost no contamination was found when labelled water was used (A). Contamination remained under 1 nL, even after 10 in-and-out pushes, although a slight increase was noted when the number of plunges increased. The contamination of the plungers was consistently greater with the solution of radiolabelled cyclophosphamide than with the pure radiolabelled solution, regardless of the test or the number of in-and-out pushes. This difference became obvious after the first use of the syringe, whether the operator touched the plunger with gloves or not; however, the total contamination of the plungers was more important after the operator had touched the plunger than otherwise, but this difference disappeared after 10 plunges.
Upper and lower surfaces of the plungers (E1 and E2). The contamination of the upper and lower surfaces of the plungers corresponds to the amount of contaminant that could come into contact with the gloves of operators.
Figure 1. Location of samples from syringe plungers.
Table 1. Amounts and volumes of cyclophosphamide on the plungers of the eight syringes
Table 2. Volumes (nL) of contaminating agents on the plungers of syringes; results of three tests